Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1172
User Manual 1178.8520.02 ─ 08
ON | 1
Switches the function on
Example:
CALC:DELT2:LINK:TO:DELT3 ON
Links D2 and D3.
CALCulate<n>:DELTamarker<ms>:LINK:TO:MARKer<md>
<State>
This command links the delta source marker <ms> to any active destination marker
<md> (normal or delta marker).
In Spectrum mode only:
●
If <md> is a delta marker, it is turned into a normal marker. To maintain the delta
marker, use
CALCulate<n>:DELTamarker<ms>:LINK:TO:DELTa<md>
●
If <md> is the reference marker for the delta marker <ms>, the delta marker is ini-
tially set to the same horizontal position as the normal marker. To change the rela-
tive distance (delta) value again, use
on page 1174. Then the delta between the two markers is maintained when you
move the normal marker.
In I/Q Analyzer mode, if <md> is the reference marker for the delta marker <ms>, the
relative distance (delta) between the two markers is maintained when you move the
normal marker.
In other applications, the delta marker is set to the same horizontal position as the
marker <md>, and if <md> is moved along the x-axis, <ms> follows to the same hori-
zontal position.
Suffix:
<n>
.
<ms>
source marker, see
<md>
destination marker, see
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:DELT4:LINK:TO:MARK2 ON
Links the delta marker 4 to the marker 2.
Example:
D2 and D3 are active.
CALC:DELT2:LINK:TO:MARK3 ON
Converts D3 into M3 and links D2 and M3
Manual operation:
See
See