
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
998
User Manual 1178.8520.02 ─ 08
Useful commands for harmonic distortion measurements described elsewhere
●
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
●
Remote commands exclusive to harmonic distortion measurements
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:BANDwidth:AUTO
............................... 998
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:NHARmonics
......................................998
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:PRESet
..............................................998
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:BANDwidth:AUTO
<State>
This command selects the resolution bandwidth of the harmonic in respect to the band-
width of the first harmonic.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
identical
ON | 1
a multiple
*RST:
1
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:NHARmonics
<NoHarmonics>
This command selects the number of harmonics that the R&S
FSV/A looks for.
Suffix:
<n>
.
<m>
Parameters:
<NoHarmonics>
Range:
1 to 26
*RST:
10
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:PRESet
This command initiates a measurement to determine the ideal configuration for the har-
monic distortion measurement.
The method depends on the span.
●
Frequency domain (span > 0)
Configuring and performing measurements