Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
261
User Manual 1178.8520.02 ─ 08
3. Select the "Save" button.
The settings are stored to a file with the extension
.xml
as specified.
How to delete an SEM settings file
1. In the "Standard Files" tab of the "Spectrum Emission Mask" dialog box, select the
file you want to delete.
2. Select the "Delete" button.
3. Confirm the message.
The settings file is removed from the R&S
FSV/A.
How to restore default SEM settings files
The R&S
FSV/A is delivered with predefined settings files which can be edited and
overwritten. However, you can restore the original files.
► In the "Standard Files" tab of the "Spectrum Emission Mask" dialog box, select the
"Restore Standard Files" button.
The original predefined settings files are available for selection on the R&S
FSV/A.
5.6.6.2
How to save SEM result files
The Result Summary from an SEM measurement can be saved to a file, which can be
exported to another application for further analysis, for example.
For details on the file format of the SEM export file, see
export format (spectrum emission mask)"
1. Configure and perform an SEM measurement as described in
to perform a spectrum emission mask measurement"
2. In the "Overview", select the "Analysis" button.
3. If necessary, change the "Decimal Separator" to "COMMA" for evaluation in other
languages.
4. Select the "Save" button.
5. In the file selection dialog box, select a storage location and filename for the result
file.
6. Select the "Save" button.
The file with the specified name and the extension
.dat
is stored in the defined
storage location.
Spectrum emission mask (SEM) measurement