![Rohde & Schwarz FSV3000 User Manual Download Page 441](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_1477843441.webp)
Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
441
User Manual 1178.8520.02 ─ 08
External noise sources are useful when you are measuring power levels that fall below
the noise floor of the R&S
FSV/A itself, for example when measuring the noise level of
an amplifier.
In this case, you can first connect an external noise source (whose noise power level is
known in advance) to the R&S
FSV/A and measure the total noise power. From this
value, you can determine the noise power of the R&S
FSV/A. Then when you measure
the power level of the actual DUT, you can deduct the known noise level from the total
power to obtain the power level of the DUT.
Remote command:
6.2.8
IF/Video output
Access
: [Input/Output] > "Output" > "IF/Video" tab
Defines the type of signal available at the "IF" and "Video" output connectors of the
R&S
FSV/A.
For restrictions and additional information, see
Chapter 6.2.1.3, "IF and video signal
IF Output State
Enables or disables output of the measured IF value at the "IF" output connector.
Remote command:
IF Out Frequency
Defines or indicates the frequency at which the IF signal level is provided at the "IF"
output connector if
is enabled.
For more information, see
Chapter 6.2.1.3, "IF and video signal output"
Remote command:
Video Output State
Enables or disables output of the displayed video signal (i.e. the filtered and detected
IF signal) at the "Video" output connector.
Remote command:
6.2.9
Audio output
Access
: [Input/Output] > "Output" > "Audio" tab
Data input and output