General instrument setup
R&S
®
FSVA3000/ R&S
®
FSV3000
688
User Manual 1178.8520.02 ─ 08
Self-alignment results in secure user mode
Be sure to store self-alignment results before
ter 3.2.15, "Protecting data using the secure user mode"
In secure user mode, the results are not stored permanently. Thus, if the currently
stored self-alignment results are not suitable, you must perform a self-alignment each
time you switch on the R&S
FSV/A.
Await Warm-Up Operation before Self Alignment
....................................................... 689
Shut down Device after Self Alignment
.......................................................................689
..........................................................................................689
Starting Touch Screen Alignment
................................................................................690
Start Self Alignment
Starts recording correction data for the instrument. If the correction data acquisition
fails or if the correction values are deactivated, a corresponding message is displayed
in the status field.
For details, see
Chapter 10.1.1, "Basics on alignment"
Note:
A running Sequencer operation is aborted when you start a self-alignment.
During self-alignment, do not connect a signal to the RF input connector. Running a
self-alignment with a signal connected to the RF input can lead to false measurement
results.
Remote command:
Alignment