Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1049
User Manual 1178.8520.02 ─ 08
Suffix:
<fe>
.
1
Connected frontend
Return values:
<Result>
string containing xml data in double quotes
For details, see
"Format of selftest result file"
Example:
EFR:SELF:RES?
See
"Example for selftest result xml file"
Usage:
Query only
Manual operation:
See
Format of selftest result file
As a result of the selftest, an XML file with test details is created. It contains the follow-
ing information in the specified order. Mandatory elements and attributes are indicated
in bold font.
Element
Attributes
Description
<Sequence>
Main element
Name
="Selftest"
Description
Optional description of the test process
FirmwareVersion
Firmware version of the controlling instrument (R&S
FSV/A)
FrontendLibrary-
Version
Version of the
control.dll
with the format
x.y.z
FrontendServerVer-
sion
Version of the RRH server with the format
x.y.z
(FE44A
only)
Date
Date the selftest was performed, with the format
dd/mm/yyyy
Time
Time the selftest was performed, with the format
hh:mm:ss
State
Test result state, combined result of all <SequenceCate-
gory>s:
"PASSED"
/
"FAILED"
Version
For internal use.
Comment
Optional comment on the test process
<SequenceCate-
gory>
Set of test steps
Name
Name of the test sequence, e.g. "Frontend voltages"
Description
Optional description of the test sequence
State
Test sequence result state, combined result of all <Sequence-
Step>s:
"PASSED"
/
"FAILED"
Version
For internal use.
Comment
Optional comment on the test sequence
Setting basic measurement parameters