Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1053
User Manual 1178.8520.02 ─ 08
<State>PASSED</State>
<Version>1.0.0</Version>
</SequenceStep>
<SequenceStep>
<Name>SUPPLY_4V</Name>
<LimitLow>3.650</LimitLow>
<LimitHigh>4.460</LimitHigh>
<MeasValue>4.053</MeasValue>
<State>PASSED</State>
<Version>1.0.0</Version>
</SequenceStep>
<SequenceStep>
<Name>SUPPLY_6V</Name>
<LimitLow>5.400</LimitLow>
<LimitHigh>6.600</LimitHigh>
<MeasValue>6.076</MeasValue>
<State>PASSED</State>
<Version>1.0.0</Version>
</SequenceStep>
<SequenceStep>
<Name>SUPPLY_M5V</Name>
<LimitLow>-6.050</LimitLow>
<LimitHigh>-4.950</LimitHigh>
<MeasValue>-5.507</MeasValue>
<State>PASSED</State>
<Version>1.0.0</Version>
</SequenceStep>
</SequenceCategory>
</Sequence>
Programming example: configuring an external frontend
The following example describes how to configure RF frontend settings in remote oper-
ation.
// Prepare the instrument
// Preset
*RST
// Create new IQ-Analyzer channel
:INST:SEL IQ
// Enable 640MHz Reference
:ROSC:O640 ON
//Enable general use of external frontend
SENSe:EFRontend:STATe ON
//Configure connection to ext. FE named "FE44S-1000826"
SENSe:EFRontend:CONNection:CONFig "FE44S","FE44S-1000826"
//Activate exclusive use of frontend by R&S
FSV/A.
SENSe:EFRontend:CONNection:STATe ON
//For demonstration purposes only: assign a static IP address
Setting basic measurement parameters