Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
858
User Manual 1178.8520.02 ─ 08
//Switch into first IQ channel to get results
INST:SEL 'IQ 1';*WAI
CALCulate:MARKer:MAXimum
CALC:MARK:Y?
//Change sweep time in IQ
SENS:SWE:TIME 300us
//Switch to single Sequencer mode
INITiate:SEQuencer:MODE SINGle
//Sweep all channels once, taking the sweep count in each channel into account
INITiate:SEQuencer:IMMediate;*OPC?
//Set marker to maximum in IQ1 and query result
CALCulate:MARKer:MAXimum
CALC:MARK:Y?
//Switch to second IQ channel and retrieve results
INST:SEL 'IQ 2';*WAI
CALCulate:MARKer:MIN
CALC:MARK:Y?
//Switch to first Spectrum channel
INST:SEL 'Spectrum';*WAI
//Query one of the SEM results
CALCulate:MARKer:FUNCtion:POWer:RESult? CPOWer
//Switch to second Spectrum channel
INST:SEL 'Spectrum 2';*WAI
//Query channel power result
CALCulate:MARKer:FUNCtion:POWer:RESult? ACPower
12.5
Configuring and performing measurements
The following commands are required to configure measurements in a remote environ-
ment. The tasks for manual operation are described in
....................................................................................859
Configuring power measurements
........................................................................ 862
Measuring the channel power and ACLR
............................................................. 866
Measuring the carrier-to-noise ratio
...................................................................... 924
Measuring the occupied bandwidth
.......................................................................925
Measuring the spectrum emission mask
...............................................................927
..............................................................................962
Analyzing statistics (APD, CCDF)
.........................................................................977
Measuring the time domain power
........................................................................987
Measuring the harmonic distortion
........................................................................997
Measuring the third order intercept point
............................................................ 1001
Measuring the AM modulation depth
.................................................................. 1004
Remote commands for EMI measurements
........................................................1006
.................................................................................1020
Configuring and performing measurements