Common analysis and display functions
R&S
®
FSVA3000/ R&S
®
FSV3000
557
User Manual 1178.8520.02 ─ 08
7.3.6
Measurement example: measuring harmonics using marker func-
tions
This measurement example describes how to measure harmonics using the provided
marker functions. Note that this task can be performed much simpler using the Har-
monic Distortion measurement (see
Chapter 5.10, "Harmonic distortion measurement"
Signal generator settings (e.g. R&S SMW):
Frequency:
128 MHz
Level:
+15 dBm
Procedure:
1. Preset the R&S
FSV/A.
2. Set the center frequency to
128 MHz
.
3. Set the span to
100 kHz
.
4. Select "Auto Level".
The R&S
FSV/A displays the reference signal with a span of 100 kHz and resolu-
tion bandwidth of 1 kHz.
5. Switch on the marker by pressing the [MKR] key.
The marker is positioned on the trace maximum.
6. Set the measured signal frequency and the measured level as reference values:
a) Press the [MKR FUNC] key
b) Press the "Reference Fixed" softkey.
The position of the marker becomes the reference point. The reference point level
is indicated by a horizontal line, the reference point frequency with a vertical line.
At the same time, the delta marker 2 is switched on.
Marker usage