Network operation and remote control
R&S
®
FSVA3000/ R&S
®
FSV3000
791
User Manual 1178.8520.02 ─ 08
Language
Comment
PXA
R&S FSEA
R&S FSEB
R&S FSEM
R&S FSEK
R&S FSP
R&S FSQ
R&S FSU
R&S FSV
Notes:
●
If you select a language other than "SCPI", the GPIB address is set to 18 if it was
20 before.
●
The Start/stop frequency, reference level and number of sweep points are adapted
to the selected instrument model.
●
For R&S FSP/FSQ/FSU emulation, HP commands are not automatically also
allowed. In this case, set
to
ON
.
●
When you switch between remote control languages, the following settings or
changes are made:
SCPI:
The instrument performs a PRESET.
8566A/B, 8568A/B, 8594E; FSEA, FSEB, FSEM; FSEK:
–
The instrument performs a PRESET.
–
The following instrument settings are changed:
Table 11-16: Instrument settings for emulation of 8566A/B, 8568A/B, 8594E; FSEA, FSEB, FSEM;
FSEK instruments
Model
# of Trace
Points
Start Freq.
Stop Freq.
Ref Level
Input Coupling
8566A/B
1001
2 GHz
22 GHz
0 dBm
AC
8568A/B
1001
0 Hz
1.5 GHz
0 dBm
AC
8560E
601
0 Hz
2.9 GHz
0 dBm
AC
8561E
601
0 Hz
6.5 GHz
0 dBm
AC
8562E
601
0 Hz
13.2 GHz
0 dBm
AC
8563E
601
0 Hz
26.5 GHz
0 dBm
AC
8564E
601
0 Hz
40 GHz
0 dBm
AC
8565E
601
0 Hz
50 GHz
0 dBm
AC
8594E
401
0 Hz
3 GHz
0 dBm
AC
FSEA
500
0 Hz
3.5 GHz
-20 dBm
AC
GPIB languages