Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
221
User Manual 1178.8520.02 ─ 08
5.5.3
OBW configuration
Access
: "Overview" > "Select Measurement" > "OBW" > "OBW Config"
This measurement is not available in zero span.
Configuring search limits for OBW measurement
The OBW measurement uses the same search limits as defined for marker search
(see
on page 527). However, only the left and right limits are consid-
ered.
Auto measurement
Use the
function to determine the required measurement parameters automat-
ically.
The following settings are determined:
●
Center frequency
●
Span: determined signal bandwidth plus 40
%
●
RBW = Span/1000
●
Sweep time = <Sweep_time>
min
*10
●
For pulsed signals: gating parameters
See also
Chapter 5.1, "Auto measurements"
Occupied bandwidth measurement (OBW)