Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
969
User Manual 1178.8520.02 ─ 08
[SENSe:]LIST:RANGe<ri>:INPut:GAIN[:VALue]
<Gain>
This command selects the preamplification level for the range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Example:
LIST:RANG2:INP:GAIN 15
[SENSe:]LIST:RANGe<ri>:LIMit:STARt
<Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Level>
Absolute limit at the start frequency of a SEM range.
Range:
-400 to 400
*RST:
13
Default unit: dBm
Example:
LIST:RANG2:LIM:STAR 200
Manual operation:
See
"Abs Limit Start/ Abs Limit Stop"
[SENSe:]LIST:RANGe<ri>:LIMit:STATe
<State>
This command turns the limit check for all spurious emission measurement ranges on
and off.
Suffix:
<ri>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
LIST:RANG2:LIM:STAT ON
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:LIMit:STOP
<Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Configuring and performing measurements