Parameters:
<Time>
Range:
125 ns to 16 s
*RST:
5 ms
Default unit: S
Example:
SWE:EGAT:CONT:PLEN 10
Manual operation:
See
[SENSe:]SWEep:EGATe:CONTinuous[:STATe]
<State>
Activates or deactivates continuous gating.
This setting is only available if
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
SWE:EGAT ON
Activate gating
SWE:EGAT:CONT:STAT ON
Activate continuous gating
Manual operation:
See
[SENSe:]SWEep:EGATe:HOLDoff
<DelayTime>
This command defines the delay time between the gate signal and the continuation of
the measurement.
Note:
If you perform gated measurements in combination with the IF Power trigger, the
R&S
FSV/A ignores the holding time for frequency sweep, FFT sweep, zero span and
I/Q mode measurements.
Parameters:
<DelayTime>
Range:
0 s to 30 s
*RST:
0 s
Default unit: S
Example:
SWE:EGAT:HOLD 100us
Manual operation:
See
[SENSe:]SWEep:EGATe:LENGth
<GateLength>
This command defines the gate length.