Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
227
User Manual 1178.8520.02 ─ 08
Label
Description
General information
"Standard"
Loaded standard settings
"Tx Power"
Power of the reference range
"Tx Bandwidth"
Tx bandwidth used by the reference range
"RBW"
RBW used by the reference range
Range results
"Range Low"
Start of the frequency range the peak value was found in
"Range Up"
Frequency range end the peak value was found in
"RBW"
RBW of the range
"Frequency"
Frequency of the peak power level
"Power Abs"
Absolute peak power level within the range
"Power Rel"
Peak power level within the range, relative to the "Tx Power"
"
Δ
Limit"
Deviation of the peak power level from the limit line
You can define in which detail the data is displayed in the result summary in the "List
Evaluation" settings (see
Chapter 5.6.5.7, "List evaluation (results configuration)"
on page 255). By default, one peak per range is displayed. However, you can change
the settings to display only peaks that exceed a threshold ("Margin").
Detected peaks are not only listed in the Result Summary, they are also indicated by
colored squares in the diagram (optionally, see
settings).
Spectrum emission mask (SEM) measurement