
General instrument setup
R&S
®
FSVA3000/ R&S
®
FSV3000
769
User Manual 1178.8520.02 ─ 08
How to use predefined parameter couplings
1. Select the [SETUP] key.
2. Select "Parameter Coupling".
3. In the "General" tab, set the parameter you want to synchronize over all measure-
ment channels to "On".
4. Close the dialog box.
How to create user-defined parameter couplings
1. Select the [SETUP] key.
2. Select "Parameter Coupling".
3. Select the "User Coupling" tab.
4. Select "Add New User Coupling".
5. From the "Channel 1" list, select the channels or type of channel to couple.
6. From the "Coupling Element 1" list, select the parameter or marker to couple for
the selected measurement channels.
To shorten the list and restrict it to a certain type of parameters, select a "Category"
first.
7. For AM/FM/PM Analog Demod channels and frequency markers only, select the
individual windows (in the frequency domain) to couple.
8. To couple specific channels rather than all channels of a type:
a) Select the second channel for coupling from the "Channel 2" list
b) Select the parameter to which the first parameter is coupled from the "Coupling
Element 2" list.
c) For AM/FM/PM Analog Demod channels and frequency markers only, select
the individual windows (in the frequency domain) from the second AM/FM/PM
Analog Demod channel to couple.
9. Select "Couple Selected Parameters".
The "Add New User Coupling" dialog box is closed, and the new user-defined cou-
pling is added to the list in the "Parameter Coupling" dialog box.
10. If specific channels are coupled, select the "Direction" to define which channel con-
trols the other, that is: in which channel the parameter is adapted if the other is
changed.
11. Close the dialog box.
From now on, if you change a coupled parameter in one channel, the parameter in
the coupled channel or channels is set to the same value.
Synchronizing measurement channel configuration