Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
146
User Manual 1178.8520.02 ─ 08
Figure 5-6: Measurement of two equally-leveled RF sinusoidal signals with the resolution band-
width which corresponds to the frequency spacing of the signals
Matching generator and R&S
FSV/A frequencies
The level drop is located exactly in the center of the display only if the generator
frequencies match the frequency display of the R&S
FSV/A exactly. To achieve
exact matching, the frequencies of the generators and the R&S
FSV/A must be
synchronized.
5. Set the resolution bandwidth to
100
kHz
.
It is no longer possible to clearly distinguish the two generator signals.
Basic measurements