Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
417
User Manual 1178.8520.02 ─ 08
Mixer Name
Specifies the name of the external mixer to which the table applies. This setting is
checked against the current mixer setting before the table can be assigned to the
range.
Remote command:
Mixer S/N
Specifies the serial number of the external mixer to which the table applies.
The specified number is checked against the currently connected mixer number before
the table can be assigned to the range.
Remote command:
[SENSe:]CORRection:CVL:SNUMber
Mixer Type
Specifies whether the external mixer to which the table applies is a two-port or three-
port type. This setting is checked against the current mixer setting before the table can
be assigned to the range.
Remote command:
Position/Value
Each position/value pair defines the conversion loss value in dB for a specific fre-
quency. The reference values must be entered in order of increasing frequencies. A
maximum of 50 reference values can be entered. To enter a new value pair, select an
empty space in the "Position"/"Value" table, or select the
button.
Correction values for frequencies between the reference values are interpolated. Lin-
ear interpolation is performed if the table contains only two values. If it contains more
than two reference values, spline interpolation is carried out. Outside the frequency
range covered by the table, the conversion loss is assumed to be the same as for the
first and last reference value.
The current configuration of the conversion loss function as described by the position/
value entries is displayed in the preview pane to the right of the table.
Remote command:
Insert Value
Inserts a new position/value entry in the table.
If the table is empty, a new entry at 0 Hz is inserted.
If entries already exist, a new entry is inserted above the selected entry. The position of
the new entry is selected such that it divides the span to the previous entry in half.
Delete Value
Deletes the currently selected position/value entry.