Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1236
User Manual 1178.8520.02 ─ 08
CALCulate<n>:LIMit<li>:LOWer[:DATA]
<LimitLinePoints>
This command defines the vertical definition points of a lower limit line.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<LimitLinePoints>
Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with
FSV/A either
adds missing values or ignores surplus values.
The unit depends on
on page 1239.
*RST:
Limit line state is OFF
Default unit: DBM
Manual operation:
See
CALCulate<n>:LIMit<li>:LOWer:MARGin
<Margin>
This command defines an area around a lower limit line where limit check violations
are still tolerated.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Margin>
numeric value
*RST:
0
Default unit: dB
Manual operation:
See
CALCulate<n>:LIMit<li>:LOWer:MODE
<Mode>
This command selects the vertical limit line scaling.
Suffix:
<n>
.
<li>
Parameters:
<Mode>
ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
Analyzing measurements (basics)