![Rohde & Schwarz FSV3000 User Manual Download Page 768](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_1477843768.webp)
General instrument setup
R&S
®
FSVA3000/ R&S
®
FSV3000
768
User Manual 1178.8520.02 ─ 08
Note that only one channel can control a generator at any time. If a measurement
channel already has control of the connected generator and you enable parameter
coupling, the control is disabled in the other channel.
Remote command:
INSTrument:COUPle:GENerator:STATe
Center Frequency
Couples the center frequency of the connected signal generator to the R&S
FSV/A.
Remote command:
INSTrument:COUPle:GENerator:CENTer[:STATe]
Generator Frequency Offset
Defines a fixed offset to the center frequency of the R&S
FSV/A for the coupled signal
generator.
Remote command:
INSTrument:COUPle:GENerator:CENTer:OFFSet
Reference level
Couples the reference level of the connected signal generator to the R&S
FSV/A.
Remote command:
INSTrument:COUPle:GENerator:RLEVel[:STATe]
Generator Reference Level Offset
Defines a fixed offset to the reference level of the R&S
FSV/A for the coupled signal
generator.
Remote command:
INSTrument:COUPle:GENerator:RLEVel:OFFSet
10.10.4
How to synchronize parameters
Access
: [SETUP] > "Parameter Coupling"
User-defined couplings allow you to couple parameters other than those available in
the "General" tab of the coupling manager. Thus, you can create highly customized
couplings between measurement channels.
Compared to the predefined couplings, user couplings do not necessarily have to syn-
chronize all active measurement channels. Instead you can define specific channels
that are synchronized with each other, in any combination, while other channels remain
independent.
Synchronizing measurement channel configuration