Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
356
User Manual 1178.8520.02 ─ 08
6.2.1.3
IF and video signal output
The measured IF signal or displayed video signal (i.e. the filtered and detected IF sig-
nal) can be provided at the IF output connector of the R&S
FSV/A.
The
video output
is a signal of 1
V. It can be used, for example, to control demodula-
ted audio frequencies.
The
IF output
is a signal of the measured level at a specified frequency.
Restrictions
Note the following restrictions for data output:
●
IF and video output is only available in the time domain (zero span).
●
For I/Q data, only IF output is available.
●
IF output is not available if any of the following conditions apply:
–
The sample rate is larger than 200 MHz (upsampling)
6.2.2
Input source settings
Access
: "Overview" > "Input" > "Input Source"
The input source determines which data the R&S
FSV/A analyzes.
The default input source for the R&S
FSV/A is "Radio Frequency", i.e. the signal at the
"RF Input" connector of the R&S
FSV/A. If no additional options are installed, this is the
only available input source.
........................................................................................... 356
6.2.2.1
Radio frequency input
Access
: "Overview" > "Input/Frontend" > "Input Source" > "Radio Frequency"