Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
860
User Manual 1178.8520.02 ─ 08
INITiate<n>:CONMeas
This command restarts a (single) measurement that has been stopped (using
ABORt
)
or finished in single sweep mode.
The measurement is restarted at the beginning, not where the previous measurement
was stopped.
Suffix:
<n>
.
irrelevant
Example:
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
INIT:CONM;*WAI
Continues the measurement (next 20 sweeps) and waits for the
end.
Result: Averaging is performed over 40 sweeps.
Manual operation:
See
INITiate<n>:CONTinuous
<State>
This command controls the sweep mode for an individual channel.
Note that in single sweep mode, you can synchronize to the end of the measurement
with *OPC, *OPC? or *WAI. In continuous sweep mode, synchronization to the end of
the measurement is not possible. Thus, it is not recommended that you use continuous
sweep mode in remote control, as results like trace data or markers are only valid after
a single sweep end synchronization.
For details on synchronization see
.
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
Continuous sweep
OFF | 0
Single sweep
*RST:
1
<State>
ON | OFF | 0 | 1
ON | 1
Continuous sweep
Configuring and performing measurements