Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1041
User Manual 1178.8520.02 ─ 08
Suffix:
<fe>
.
1
Connected frontend
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
The connection to the frontend is deactivated temporarily. The
frontend is thus available for use elsewhere, for example by a
signal generator. The measurement settings on the R&S
FSV/A
remain untouched.
ON | 1
Frontend connection enabled.
The frontend is reserved for exclusive use by the R&S
FSV/A.
*RST:
0
Example:
//Global activation of external frontend
EFR ON
//Configure frontend
EFR:CONN:CONF "FE44S","123.456.789"
//Activate exclusive use of frontend by
R&S FSV/A.
EFR:CONN ON
Manual operation:
See
"External Frontend Connection State"
[SENSe:]EFRontend<fe>:CONNection:CONFig
<Type>, <IPAddress>[, <DeviceID>,
<SymbolicName>]
Configures the connection to the external frontend.
Suffix:
<fe>
.
1
Connected frontend
Parameters:
<Type>
"FE44S" | "FE50DTR"
String in double quotes containing the type of frontend to be
connected.
<IPAddress>
string in double quotes
The IP address or computer name of the frontend connected to
the R&S
FSV/A via LAN. The IP address and computer name
are indicated on the electronic ink display on the side panel of
the frontend.
<DeviceID>
string in double quotes
Unique device ID consisting of <type>-<serialnumber>
Not required or relevant for the R&S
FSV/A.
Setting basic measurement parameters