![Rohde & Schwarz FSV3000 User Manual Download Page 1185](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_14778431185.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1185
User Manual 1178.8520.02 ─ 08
Suffix:
<n>
.
<m>
Manual operation:
See
CALCulate<n>:MARKer<m>:MINimum:AUTO
<State>
This command turns an automatic marker peak search for a trace minimum on and off.
The R&S
FSV/A performs the peak search after each sweep.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:MIN:AUTO ON
Activates the automatic minimum value search function for
marker 1 at the end of each particular sweep.
Manual operation:
See
"Auto Max Peak Search / Auto Min Peak Search"
CALCulate<n>:MARKer<m>:MINimum:LEFT
This command moves a marker to the next minimum peak value.
The search includes only measurement values to the right of the current marker posi-
tion.
In the spectrogram, the command moves a marker horizontally to the minimum level in
the currently selected frame. The vertical marker position remains the same.
Suffix:
<n>
.
<m>
Manual operation:
See
CALCulate<n>:MARKer<m>:MINimum:NEXT
This command moves a marker to the next minimum peak value.
In the spectrogram, the command moves a marker horizontally to the minimum level in
the currently selected frame. The vertical marker position remains the same.
Analyzing measurements (basics)