Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
438
User Manual 1178.8520.02 ─ 08
Status bar message
Dialog status
SYST:ERR? result
Description/ Possible solution
External Frontend [x]:
LAN Connection Failed
Not connected
301, "LAN Connection
Failed"
Verify that the external frontend is connected to the
same network as the R&S
FSV/A.
Verify the specified IP address of the external fron-
tend.
External Frontend [x]:
Incompatible Analyzer
Firmware
Incompatible Ana-
lyzer Firmware
302, "Incompatible
Analyzer Firmware"
The frontend hardware is not compatible with the
installed analyzer firmware.
Update your analyzer firmware.
Chapter 10.8.4, "Firmware updates"
External Frontend [x]:
Incompatible Frontend
Firmware
Incompatible Fron-
tend Firmware
303, "Incompatible
Frontend Firmware"
The firmware of the external frontend is not up-to-
date.
Perform a firmware update.
External Frontend [x]:
Device Locked
Device Locked
304, "Device Locked
"
External frontend is accessible via ping, but no
SSL-connection could be established.
Disconnect another device that is currently access-
ing the external frontend and try reconnecting.
A single frontend supports up to 2 LAN/SSL con-
nections.
External Frontend [x]:
Wrong Device Type
Wrong Device Type
306, "Wrong Device
Type"
The connected external frontend is not of the
specified type.
Disconnect the device, select the correct external
frontend type in the configuration dialog box, and
reconnect.
External Frontend [x]:
LAN Connection Lost
LAN Connection Lost
314, "LAN Connection
Lost"
The connection to the external frontend has been
lost.
The analyzer attempts to reconnect periodically.
Wait for the "Connecting…" message.
External Frontend [x]:
Firmware Update Failed
Firmware Update
Failed
315, "Firmware
Update Failed"
The firmware of the external frontend was not
updated successfully.
External Frontend [x]:
Settings Mismatch -
reinitiate frontend by
changing a measure-
ment parameter
Connected
332, "Settings
Mismatch"
The conversion settings on the external frontend
were changed by another instrument during the
measurement on the R&S
FSV/A. The measure-
ment results no longer match the original measure-
ment settings. To restore the original measurement
settings on the frontend, change a frontend setting
in the R&S
FSV/A dialog.
External Frontend [x]:
Not Connected – Device
Busy
Device busy
333, "Device Busy"
The external frontend is currently blocked by
another instrument, e.g. for initial connection or a
firmware update. After the other instrument finishes
its task, try reconnecting the R&S
FSV/A. (Switch
the "Connection" state off and on again.)
External Frontend [x]:
Device Busy
Connected
333, "Device Busy"
During a running measurement, the external fron-
tend is blocked by another instrument, e.g. for ini-
tial connection or a firmware update. The current
measurement is aborted. After the other instrument
finishes its task, start a new measurement on the
R&S
FSV/A.
Data input and output