Network operation and remote control
R&S
®
FSVA3000/ R&S
®
FSV3000
804
User Manual 1178.8520.02 ─ 08
5. Select "Save".
A script with the required header data for the selected format is stored to a file.
11.5.5
Example for a recorded SCPI script
The following example shows a recorded SCPI script in python format. The script con-
figures the ACLR measurement example described in
example 1 – ACPR measurement on a CDMA2000 signal"
# python script created: 29:11:2017 09:08:53
import visa
# connect to instrument (MU717225)
VisaResourceManager = visa.ResourceManager()
FSV3000 = VisaResourceManager.open_resource("TCPIP::10.124.0.195::INSTR")
# Display update ON
FSV3000.write("SYST:DISP:UPD ON")
FSV3000.write("*RST")
FSV3000.write("INIT:CONT OFF")
FSV3000.write("SENS:FREQ:CENT 850000000")
FSV3000.write("SENS:FREQ:SPAN 4000000")
FSV3000.write("DISP:WIND:SUBW:TRAC:Y:SCAL:RLEV 10")
FSV3000.write("CALC:MARK:FUNC:POW:SEL ACP")
FSV3000.write("CALC:MARK:FUNC:POW:PRES S2CD")
FSV3000.write("SENS:ADJ:LEV;*WAI")
FSV3000.write("SENS:POW:HSP ON")
FSV3000.write("TRAC1:DATA? TRACE1")
FSV3000.query("CALC1:MARK:Y?")
FSV3000.write("CALC1:MARK:FUNC:POW:RES? CPOW")
# Back to local mode
FSV3000.write("@LOC")
11.6
Network and remote control settings
Access
: [SETUP] > "N Remote"
Network settings in secure user mode
Be sure to store all network settings before
is enabled; see
ter 3.2.15, "Protecting data using the secure user mode"
If the currently stored network settings are not suitable, you must correct them each
time you switch on the R&S
FSV/A in secure user mode, as the settings are not stored
permanently in this case.
Network and remote control settings