
Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
328
User Manual 1178.8520.02 ─ 08
vidual requirements on the dwell time are described for each detector. For details on
defining the dwell time for an EMI measurement, see
"Defining a dwell time for the final
Positive and negative peak detector
The maximum and minimum peak detectors display the maximum and minimum signal
level that was detected during the specified dwell time.
The minimum and maximum peak detectors are already available with the base unit.
Consider the following when defining the dwell time:
●
Unmodulated signals
: minimum time required by the detector
●
Pulsed signals
: the time must be long enough to capture at least one complete
pulse
Average detector
The average detector displays the average signal level of the samples that were collec-
ted during the specified dwell time.
The average detector is already available with the base unit.
Consider the following when defining the dwell time:
●
Unmodulated signals
: minimum time required by the detector
●
Pulsed signals
: the time must be long enough to capture several complete pulses
(at least 10)
●
The time is determined by the lowest modulation frequency to be averaged
RMS detector
The RMS detector displays the root mean square (RMS) value over the specified dwell
time. The integration time is the specified dwell time.
The RMS detector is already available with the base unit.
The same considerations apply to the dwell time as for the average detector.
Sample detector
The sample detector displays the last value from the samples allocated to a pixel.
The sample detector is used for noise or phase noise marker calculation. However, it is
unreliable if the displayed span is much greater then the resolution bandwidth or if the
tuning steps of the local oscillator are too large. The sample detector is not recommen-
ded for EMI tests.
Quasi-peak detector (CISPR filter only)
The quasi-peak detector displays the maximum signal level weighted to CISPR
16-1-1
that was detected during the dwell time.
The quasi-peak detector is only available for the CISPR filter, and not for an RBW of
1
MHz.
Electromagnetic interference (EMI) measurement