Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
907
User Manual 1178.8520.02 ─ 08
Example:
SENS:POW:ACH:FILT:GAB:MAN:LOW BC,ON
Example:
See
"Example: configuring and performing an asymmetrical
Manual operation:
See
[SENSe:]POWer:ACHannel:FILTer[:STATe]:GAP<gap>:MANual:UPPer
<SBGaps>,
<State>
[SENSe:]POWer:ACHannel:FILTer[:STATe]:GAP<gap>:MANual:UPPer?
<SBGaps>
This command turns the weighting filter for the specified upper gap channel on and off.
Suffix:
<gap>
.
1 | 2
Gap channel number
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Parameters for setting and query:
<SBGaps>
AB | BC | CD | DE | EF | FG | GH
Name of the gap, defined by the letters of the surrounding sub
blocks (e.g. "AB" for the gap between sub blocks A and B).
Example:
SENS:POW:ACH:FILT:GAP:MAN:UPP BC,ON
Example:
See
"Example: configuring and performing an asymmetrical
Manual operation:
See
[SENSe:]POWer:ACHannel:FILTer:ALPHa:GAP<gap>:MANual:LOWer
<SBGaps>,
<Alpha>
[SENSe:]POWer:ACHannel:FILTer:ALPHa:GAP<gap>:MANual:LOWer?
<SBGaps>
This command defines the roll-off factor for the specified lower gap channel's weighting
filter.
Suffix:
<gap>
.
1 | 2
Gap channel number
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Configuring and performing measurements