Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
239
User Manual 1178.8520.02 ─ 08
sured values are then compared with this single limit line. If the limit is exceeded in any
range, the result of the limit check is ."' failed!"
Figure 5-44: Summarized limit line for multiple sub blocks
5.6.5
SEM configuration
Access
: "Overview" > "Select Measurement" > "Spectrum Emission Mask"
The SEM measurement is started immediately with the default settings.
Auto measurement
Use the
function to determine the required measurement parameters automat-
ically.
The following settings are determined:
●
Center frequency
●
Span: determined signal bandwidth plus 40
%
●
SEM standard according to estimated signal bandwidth
●
For pulsed signals: gating parameters
After the parameters have been determined automatically, you must confirm the SEM
standard selection.
See also
Chapter 5.1, "Auto measurements"
The remote commands required to perform these tasks are described in
ter 12.5.6, "Measuring the spectrum emission mask"
Spectrum emission mask (SEM) measurement