Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1122
User Manual 1178.8520.02 ─ 08
INPut<ip>:GAIN[:VALue]
<Gain>
This command selects the "gain" if the preamplifier is activated (
INP:GAIN:STAT ON
,
The command requires the additional preamplifier hardware option.
For R&S
FSV/A44 or higher models, note the restrictions described in
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Gain>
For R&S FSV/A3004, 3007, 3013, and 3030 models, the follow-
ing settings are available:
15 dB and 30 dB
All other values are rounded to the nearest of these two.
R&S
FSV/A44 or higher:
30 dB
Default unit: DB
Example:
INP:GAIN:STAT ON
INP:GAIN:VAL 30
Switches on 30 dB preamplification.
Manual operation:
See
12.7.4.4
Scaling the Y-axis
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]
.......................................1122
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE
............................1123
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:PDIVision
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RPOSition
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
......................................1124
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]
<Range>
This command defines the display range of the y-axis (for all traces).
Note that the command works only for a logarithmic scaling. You can select the scaling
with
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
.
Suffix:
<n>
.
<w>
subwindow
Not supported by all applications
<t>
irrelevant
Setting basic measurement parameters