Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
954
User Manual 1178.8520.02 ─ 08
[SENSe:]ESPectrum<sb>:MSR:CLASs
<Class>
Defines the class of the base station according to its sending range.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
Parameters:
<Class>
WIDE | MEDium | LOCal
WIDE
Wide Area
MEDium
Medium Range
LOCal
Local Area
*RST:
WIDE
Example:
//Select the band category 1
ESP2:MSR:BCAT 1
//Set the base station class to medium range
ESP2:MSR:CLAS MED
//Set the maximum output power to 10 dBm.
ESP2:MSR:MPOW 10
//Set the frequency range of the base station to > 3 GHz
ESP2:MSR:BAND:HIGH
//Set the base station RF bandwidth to 20 MHz
ESP2:MSR:RFBW 20MHZ
//Calculate limits for MSR SEM
ESP2:MSR:APPL
Manual operation:
See
[SENSe:]ESPectrum<sb>:MSR:GSM:CARRier
<Power>
Defines the power of the GSM carrier (if available, see
This command is only available for band category 2 (see
Suffix:
<sb>
.
1 to 8
Parameters:
<Power>
Range:
0 dBm to 100 dBm
*RST:
39.0 dBm
Default unit: dBm
Configuring and performing measurements