Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1084
User Manual 1178.8520.02 ─ 08
Manual operation:
See
[SENSe:]CORRection:TRANsducer:GENerate
<Name>
This command uses the normalized measurement data to generate a transducer factor
with up to 1001 points. The trace data is converted to a transducer with unit dB and
stored in a file with the specified name and the suffix
.trd
under
C:\Program Files\Rohde-Schwarz\FSV3000\<version>\trd
. The frequency
points are allocated in equidistant steps between start and stop frequency.
Parameters:
<Name>
'<name>'
Example:
CORR:TRAN:GEN 'MyGenerator'
Creates the transducer file
C:\r_s\instr\trd\MyGenerator.trd
.
Manual operation:
See
Programming example for external generator control
The following example demonstrates how to work with an external generator in a
remote environment.
It assumes that a signal generator of the type SMA01A is connected to the R&S
FSV/A
via TCP/IP, as described in
"External generator connections"
//--------------Preparing the instrument -----------
//Reset the instrument
*RST
//Set the frequency span.
SENS:FREQ:STAR 10HZ
SENS:FREQ:STOP 1MHZ
//--------------Configuring the interface -----------
//Set the generator type to SMW06 with a frequency range of 100 kHz to 3GHz
SYST:COMM:RDEV:GEN:TYPE 'SMA01A'
//Set the interface used to the TCP/IP address 130.094.122.195
SYST:COMM:RDEV:GEN:INT TCP
SYST:COMM:TCP:RDEV:GEN:ADDR 130.094.122.195
//Activate the use of the external reference frequency at 10 MHz on the generator
SOUR:EXT:ROSC EXT
//--------------Configuring the calibration measurement -----------
//Activate external generator control.
SOUR:EXT:STAT ON
Setting basic measurement parameters