Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1056
User Manual 1178.8520.02 ─ 08
Parameters:
<BiasSetting>
*RST:
0.0 A
Default unit: A
Manual operation:
See
[SENSe:]MIXer<x>:BIAS[:LOW]
<BiasSetting>
This command defines the bias current for the low (first) range.
This command is only available if the external mixer is active (see
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<BiasSetting>
*RST:
0.0 A
Default unit: A
Manual operation:
See
[SENSe:]MIXer<x>:LOPower
<Level>
This command specifies the LO level of the external mixer's LO port.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<Level>
numeric value
Range:
13.0 dBm to 17.0 dBm
Increment: 0.1 dB
*RST:
15.5 dBm
Example:
MIX:LOP 16.0dBm
Manual operation:
See
[SENSe:]MIXer<x>:SIGNal
<State>
This command specifies whether automatic signal detection is active or not.
Note that automatic signal identification is only available for measurements that per-
form frequency sweeps (not in vector signal analysis or the I/Q Analyzer, for instance).
The "Auto ID" function is now also available for
Spectrum emission mask (SEM) mea-
Spurious emissions measurement
Setting basic measurement parameters