Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1088
User Manual 1178.8520.02 ─ 08
[SENSe:]PMETer<p>:UPDate[:STATe]
............................................................................1093
........................................................................................1094
.............................................................................. 1094
CALibration:PMETer<p>:ZERO:AUTO ONCE
This command zeroes the power sensor.
Note that you have to disconnect the signals from the power sensor input before you
start to zero the power sensor. Otherwise, results are invalid.
Suffix:
<p>
.
Power sensor index
Example:
CAL:PMET2:ZERO:AUTO ONCE;*WAI
Starts zeroing the power sensor 2 and delays the execution of
further commands until zeroing is concluded.
Usage:
Event
Manual operation:
See
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]
<RefValue>
This command defines the reference value for relative measurements.
Suffix:
<n>
.
<p>
Power sensor index
Parameters:
<RefValue>
Range:
-200 dBm to 200 dBm
*RST:
0
Default unit: DBM
Example:
CALC:PMET2:REL -30
Sets the reference value for relative measurements to -30 dBm
for power sensor 2.
Manual operation:
See
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
This command sets the current measurement result as the reference level for relative
measurements.
Suffix:
<n>
.
<p>
Power sensor index
Example:
CALC:PMET2:REL:AUTO ONCE
Takes the current measurement value as reference value for rel-
ative measurements for power sensor 2.
Setting basic measurement parameters