Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
269
User Manual 1178.8520.02 ─ 08
Child node
Attribute
Value
Parameter description
Mand.
Unit
dBm
Defines dBm as unit
Yes, if the Refer-
enceLevel node
is used
RFAttenuation
Mode
Manual | Auto
Yes, if the Refer-
enceLevel child
node is used
Preamplifier
ON | OFF | 1 | 0
Yes
5.6.8.2
ASCII file export format (spectrum emission mask)
When trace data from an SEM measurement is exported, the data is stored in ASCII
format as described below. The first part of the file lists information about the signal
analyzer and the general setup.
File contents
Explanation
File header
Type;FSVA3-26;
Model
Version;1.00;
Firmware version
Date;31.Mar 17;
Storage date of data set
Mode;ANALYZER;SEM;
Operating mode and measurement function
Center Freq;13250000000.000000;Hz
X-axis settings
Freq Offset;0.000000;Hz
Span;25500000.000000;Hz
x-Axis;LIN;
Start;13237250000.000000;Hz
Stop;13262750000.000000;Hz
Level Offset;0.000000;dB
Y-axis settings
Ref Position;100.000000;%
y-Axis;LOG;
Level Range;100.000000;dB
Trace settings
Trace Mode;CLR/WRITE;
Detector;RMS;
Sweep Count;0;
Trace 1:;
x-Unit;Hz;
y-Unit;dBm;
Spectrum emission mask (SEM) measurement