![Rohde & Schwarz FSV3000 User Manual Download Page 1358](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_14778431358.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1358
User Manual 1178.8520.02 ─ 08
INSTrument:COUPle:USER<uc>:CHANnel:LIST?
This command queries the names of the measurement channels that can be synchron-
ized.
Suffix:
<uc>
.
irrelevant
Return values:
<SynchronizableChannel>
Comma-separated list of strings
All channels that can be synchronized.
Example:
INST:COUP:USER:CHAN:LIST?
Result:
'SPEC1','AD1','All Spectrum','All
Channels','All Analog Demod'
Usage:
Query only
Manual operation:
See
INSTrument:COUPle:USER<uc>:ELEMent:LIST?
[<ChannelName>, <Parameter>]
This command queries parameters that can be synchronized.
Suffix:
<uc>
.
irrelevant
Query parameters:
<ChannelName>
Optional SCPI parameter.
String containing the name of a measurement channel.
<Parameter>
Optional SCPI parameter.
String containing the name of a parameter that you can syn-
chronize.
Return values:
<SynchronizableParameter>
Comma-separated list of parameters.
No parameters provided
Parameters that can be synchronized for all channels
Channel name provided
Parameters that can be synchronized for the selected channel
Parameter and channel name provided
Parameters that can be synchronized with the specified parame-
ter for the selected channel.
Example:
INST:COUP:USER:ELEM:LIST?
Result: all parameters that can be coupled:
'AC DC Coupling', 'Attenuation', 'Center
Frequency', 'Display Lines', 'Frequency Marker
1',...
Configuring the R&S
FSV/A