In order to recognize when the acquisition of correction data is completed, the MAV bit
in the status byte can be used. If the associated bit is set in the Service Request
Enable (SRE) register, the instrument generates a service request after the acquisition
of correction data has been completed.
Return values:
<CalibrationFailed>
ON | OFF | 0 | 1
OFF | 0
Calibration was successful.
ON | 1
Calibration was not successful.
Example:
*CLS
Resets the status management.
*SRE 16
Enables MAV bit in the Service Request Enable register.
*CAL?
Starts the correction data recording, and then a service request
is generated.
Usage:
Query only
Manual operation:
See
CALibration:RESult?
This command returns the results collected during calibration.
Return values:
<CalibrationData>
String containing the calibration data.
Example:
CAL:RES?
would return, e.g.
Total Calibration Status:
PASSED, Date (dd/mm/yyyy): 12/07/2004,
Time: 16:24:54,Runtime: 00.06
Usage:
Query only
Manual operation:
See
DIAGnostic:SERVice:INPut:MC:CFRequency
<Frequency>
This command defines the frequency of the calibration signal for R&S
FSV/A models
13 GHz and higher.
This command only takes effect if a microwave calibration signal is selected for input
(
DIAGnostic:SERVice:INPut[:SELect]
Parameters:
<Frequency>
*RST:
8.004 GHz
Default unit: Hz
Manual operation:
See
Configuring the R&S
FSV/A