Getting started
R&S
®
FSVA3000/ R&S
®
FSV3000
55
User Manual 1178.8520.02 ─ 08
A message in the status bar ("Instrument warming up...") indicates that the operating
temperature has not yet been reached.
Depending on the installation settings, an automatic self-alignment is performed each
time the instrument is switched on. A dialog box is displayed indicating how much
warm-up time is still required before self-alignment can be performed.
1. Press the [Setup] key.
2. Press the "Alignment" softkey.
3. Select the "Start Self Alignment" button in the "Alignment" dialog box.
Once the system correction values have been calculated successfully, a message
is displayed.
To display the alignment results again later
●
Press the [SETUP] key.
●
Press the "Alignment" softkey.
3.2.14
Considerations for test setup
Cable selection and electromagnetic interference (EMI)
Electromagnetic interference (EMI) can affect the measurement results.
To suppress electromagnetic radiation during operation:
●
Use high-quality shielded cables, for example, double-shielded RF and LAN
cables.
●
Always terminate open cable ends.
●
Ensure that connected external devices comply with EMC regulations.
Signal input and output levels
Information on signal levels is provided in the data sheet and on the instrument, next to
the connector. Keep the signal levels within the specified ranges to avoid damage to
the R&S
FSV/A and connected devices.
3.2.15
Protecting data using the secure user mode
During normal operation, the R&S
FSV/A uses a solid-state drive to store its operating
system, instrument firmware, instrument self-alignment data, and any user data cre-
ated during operation.
Optionally, a removable solid-state drive is available (R&S
FSV3-B20), which can be
removed from the R&S
FSV/A and locked in a secure place to protect any classified
data it may contain.
Preparing for use