Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
293
User Manual 1178.8520.02 ─ 08
Remote command:
CALCulate<n>:STATistics:SCALe:X:RANGe
Shifting the Display (Offset) ← X-Axis
Defines an arithmetic level offset. This offset is added to the measured level irrespec-
tive of the selected unit. The scaling of the x-axis is changed accordingly. The setting
range is ±200 dB in 0.1 dB steps.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RLEVel:
OFFSet
Y-Axis
Defines the scaling settings for the probability distribution.
Y-Unit ← Y-Axis
Defines the scaling type of the y-axis as either percentage or absolute. The default
value is absolute scaling.
Remote command:
CALCulate<n>:STATistics:SCALe:Y:UNIT
Y-Max/ Y-Min ← Y-Axis
Defines the upper (max) and lower (min) limit of the displayed probability range. Values
on the y-axis are normalized which means that the maximum value is 1.0. The mini-
mum value must be in the range:
1E-9
<
Y-Min
<
0.1
The distance between "Y-Max" and "Y-Min" must be at least one decade.
Remote command:
CALCulate<n>:STATistics:SCALe:Y:UPPer
CALCulate<n>:STATistics:SCALe:Y:LOWer
Default Settings
Resets the x- and y-axis scalings to their preset values.
X-axis ref level:
-10 dBm
X-axis range APD:
100 dB
X-axis range CCDF:
20 dB
Y-axis upper limit:
1.0
Y-axis lower limit:
1E-6
Remote command:
CALCulate<n>:STATistics:PRESet
Statistical measurements (APD, CCDF)