Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1246
User Manual 1178.8520.02 ─ 08
Suffix:
<n>
.
<li>
irrelevant
Example:
CALC:LIM:CLE
Deletes the result of the limit check.
CALCulate<n>:LIMit<li>:FAIL?
This command queries the result of a limit check in the specified window.
Note that for SEM measurements, the limit line suffix <li> is irrelevant, as only one spe-
cific SEM limit line is checked for the currently relevant power class.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<li>
Return values:
<Result>
0
PASS
1
FAIL
Example:
INIT;*WAI
Starts a new sweep and waits for its end.
CALC2:LIM3:FAIL?
Queries the result of the check for limit line 3 in window 2.
Usage:
Query only
Manual operation:
See
See
12.8.5.4
Programming example: using limit lines
The following examples demonstrate how to work with limit lines in a remote environ-
ment.
Example: configuring limit lines
...........................................................................1246
Example: performing a limit check
...................................................................... 1248
Example: configuring limit lines
This example demonstrates how to configure 2 limit lines - an upper and a lower limit -
for a measurement in a remote environment.