Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1133
User Manual 1178.8520.02 ─ 08
Parameters:
<GateLength>
Range:
125 ns to 30 s
*RST:
400μs
Default unit: S
Example:
SWE:EGAT:LENG 10ms
Manual operation:
See
[SENSe:]SWEep:EGATe:POLarity
<Polarity>
This command selects the polarity of an external gate signal.
The setting applies both to the edge of an edge-triggered signal and the level of a
level-triggered signal.
Parameters:
<Polarity>
POSitive | NEGative
*RST:
POSitive
Example:
SWE:EGAT:POL POS
Manual operation:
See
[SENSe:]SWEep:EGATe:SOURce
<Source>
This command selects the signal source for gated measurements.
If an IF power signal is used, the gate is opened as soon as a signal at > -20 dBm is
detected within the IF path bandwidth (10 MHz).
For more information see
Parameters:
<Source>
*RST:
IFPower
Example:
SWE:EGAT:SOUR IFP
Switches the gate source to IF power.
Manual operation:
See
See
See
See
See
See
Setting basic measurement parameters