Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1071
User Manual 1178.8520.02 ─ 08
Example:
//Query part number
PROB3:ID:PART?
Usage:
Query only
Manual operation:
See
[SENSe:]PROBe<pb>:ID:SRNumber?
Queries the serial number of the probe.
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Return values:
<SerialNo>
Example:
//Query serial number
PROB3:ID:SRN?
Usage:
Query only
Manual operation:
See
[SENSe:]PROBe<pb>:SETup:ATTRatio
<AttenuationRatio>
Defines the attenuation applied to the input at the probe. This setting is only available
for modular probes.
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<AttenuationRatio>
10
Attenuation by 20 dB (ratio= 10:1)
2
Attenuation by 6 dB (ratio= 2:1)
*RST:
10
Default unit: DB
Manual operation:
See
[SENSe:]PROBe<pb>:SETup:CMOFfset
<CMOffset>
Sets the common mode offset. The setting is only available if a differential probe in
CM-mode is connected to the R&S
FSV/A.
If the probe is disconnected, the common mode offset of the probe is reset to 0.0
V.