Common analysis and display functions
R&S
®
FSVA3000/ R&S
®
FSV3000
549
User Manual 1178.8520.02 ─ 08
Marker Demodulation State
Activates or deactivates the demodulation output.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe]
Continuous Demodulation
If activated, the signal is demodulated continuously and sent to the audio output,
instead of stopping the sweep at the marker frequency of marker 1 and demodulating
there for the configured marker stop time. This allows you to monitor the frequency
range acoustically (assuming the sweep time is long enough).
For zero span measurements, demodulation is always active continuously.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous
Marker Stop Time
Defines how long the sweep is stopped at the marker position to output the demodula-
ted signal.
For zero span measurements, demodulation is always active continuously, regardless
of the marker stop time.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff
Modulation
Defines the demodulation mode for output. The default setting is AM.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect
Squelch
Activates or deactivates the squelch function. If activated, the audible AF is cut off
below a defined threshold level. Thus, you avoid hearing noise at the audio output
when no signal is available.
The squelch function activates the video trigger function (see
on page 487) and
deactivates any other trigger or gating settings. The squelch level and trigger level are
set to the same value. The trigger source in the channel bar is indicated as "SQL" for
squelch. The squelch level is indicated by a red line in the diagram.
The trigger source in the channel bar is indicated as "SQL" for squelch. The squelch
level is indicated by a red line in the diagram.
Remote command:
Marker usage