![Rohde & Schwarz FSV3000 User Manual Download Page 302](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_1477843302.webp)
Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
302
User Manual 1178.8520.02 ─ 08
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
Left Limit / Right Limit
Defines a power level limit for line S1 (left) or S2 (right).
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
5.9.5
How to measure powers in the time domain
The step-by-step procedure to measure powers in the time domain is described here in
detail.
For remote operation, see
Chapter 12.5.9.4, "Programming example: time domain
To measure the power in the time domain
1. Select the [MEAS] key.
2. From the "Select Measurement" dialog box, select the "Time Domain Power" mea-
surement function.
3. Select the type of power measurement results to be determined by selecting the
corresponding softkeys.
4. To restrict the power evaluation range, define limits:
a) Select the "Time Dom Power Config" softkey to display the "Time Domain
Power" configuration dialog box.
b) Switch on the limits by setting the "Limit State" to "On".
The limit lines S1 and S2 are displayed.
c) Define the left limit (limit line S1), the right limit (S2), or both.
5. Start a sweep.
The measured powers are displayed in the marker results.
5.9.6
Measurement example
This measurement example demonstrates the time domain power calculation for a
GSM burst.
Time domain power measurement