Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1374
User Manual 1178.8520.02 ─ 08
Setting up instrument emulation
..........................................................................1374
Reference: GPIB commands of emulated HP models
........................................1377
Reference: command set of emulated PSA models
........................................... 1405
Reference: command set of emulated PXA models
........................................... 1409
12.13.1
Setting up instrument emulation
The following commands are required to set up the use of commands to emulate other
instruments.
Useful commands for emulating other instruments described elsewhere:
●
Remote commands exclusive to emulating other instruments:
.............................................................................................. 1376
.......................................................................................... 1376
SYSTem:HPCoupling
<CouplingType>
Controls the default coupling ratios in the HP emulation mode for:
●
span and resolution bandwidth (Span/RBW) and
●
resolution bandwidth and video bandwidth (RBW/VBW)
This command is only available if a HP language is selected using
Parameters:
<CouplingType>
HP | FSP
*RST:
FSP
Example:
SYSTem:HPC HP
Manual operation:
See
SYSTem:IFGain:MODE
<Mode>
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300 kHz and is
only available if a HP language is selected using
Parameters:
<Mode>
NORMal | PULSe
Emulating other instruments' commands