Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
341
User Manual 1178.8520.02 ─ 08
●
For gated measurements, the maximum RBW is restricted to 3
MHz. If the current
RBW is larger when you enable gating, the RBW is automatically reduced to
3
●
For EMI measurements using the quasipeak detector, the 1
MHz RBW filter is not
Chapter 5.13.3.1, "Resolution bandwidth and filter types"
Remote command:
[SENSe:]BANDwidth[:RESolution]
[SENSe:]BANDwidth[:RESolution]:AUTO
Automatic Peak Search
If activated, a peak search is performed automatically for all active markers after each
sweep.
If Auto peak search
and
limit lines are active, the active markers are set to the peak
delta
values between the measured signal and the limit lines.
Note:
Auto Max Peak Search / Auto Min Peak Search
are not available for EMI measurements.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:PEAKsearch:AUTO
Dwell Time
Sets the dwell time for the EMI marker measurement.
For more information see
Chapter 5.13.3.2, "Detectors and dwell time"
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DWELl
Final Test Detector
Defines the detector to be used for the final EMI test at the marker frequency.
This setting is only available if the EMI (R&S
FSV3-K54) measurement option is instal-
led. CISPR detectors are only available after selecting a CISPR RBW or filter (see
For details, see
Chapter 5.13.3.2, "Detectors and dwell time"
Note:
The trace detector configured in the trace settings is used for the initial peak
search only, see
Chapter 7.5.1.2, "Trace settings"
"Off"
No final test is performed.
"PositivePeak"
Determines the maximum signal level that was detected during the
specified dwell time.
"Average"
Determines the average signal level of the samples that were collec-
ted during the specified dwell time.
"Quasi-Peak"
Determines the maximum signal level weighted to CISPR
16-1-1 that
was detected during the dwell time.
The "Quasi-Peak" detector is only available for the CISPR filter, and
not for an RBW of 1
MHz.
Electromagnetic interference (EMI) measurement