Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1042
User Manual 1178.8520.02 ─ 08
<SymbolicName>
string in double quotes
Symbolic name of the external frontend.
Not required or relevant for the R&S
FSV/A.
Example:
//Global activation of external frontend
EFR ON
//Configure frontend
EFR:CONN:CONF "FE44S","123.456.789"
//Activate exclusive use of frontend by
R&S FSV/A.
EFR:CONN ON
Manual operation:
See
See
[SENSe:]EFRontend<fe>:CONNection:CSTate?
Queries the status of the physical connection to the external frontend.
Suffix:
<fe>
.
1
Connected frontend
Return values:
<State>
ON | OFF | 0 | 1
OFF | 0
Frontend not connected; connection error
ON | 1
Frontend connected
Usage:
Query only
Manual operation:
See
"External Frontend Connection State"
[SENSe:]EFRontend<fe>:FREQuency:BAND:COUNt?
Queries the number of frequency bands provided by the selected frontend.
Suffix:
<fe>
.
1
Connected frontend
Return values:
<NoBands>
integer
Number of frequency bands
Example:
//Query number of frequency bands
EFR:FREQ:BAND:COUN?
//Result: 2
Usage:
Query only
Manual operation:
See
Setting basic measurement parameters