Getting started
R&S
®
FSVA3000/ R&S
®
FSV3000
71
User Manual 1178.8520.02 ─ 08
The serial number is used to define the
default instrument name
, which is:
<Type><variant>-<serial_number>
For example,
FSV3004-123456
.
The instrument name is required to establish a connection to the instrument in a LAN.
3.4
Trying out the instrument
This chapter introduces the most important functions and settings of the R&S
FSV/A
step by step. The complete description of the functionality and its usage is given in the
R&S
FSV/A User Manual. Basic instrument operation is described in
Prerequisites
●
The instrument is set up, connected to the mains system, and started up as descri-
bed in
Chapter 3.2, "Preparing for use"
For these first measurements, you use the internal calibration signal, so you do not
need any additional signal source or instruments. Try out the following:
.........................................................................................71
........................................................................................73
Activating additional measurement channels
..........................................................75
Performing sequential measurements
.................................................................... 80
.................................................................................. 81
.................................................................................. 83
......................................................................................... 84
Zooming into the display permanently
.................................................................... 87
......................................................................................92
3.4.1
Measuring a basic signal
We will start out by measuring a basic signal, using the internal calibration signal as
the input.
To display the internal 64 MHz calibration signal
1. Press the [PRESET] key to start out in a defined instrument configuration.
2. Press the [Setup] key on the front panel.
3. Tap the "S Support" softkey.
4. Tap the "Calibration Signal" tab.
Trying out the instrument