Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1060
User Manual 1178.8520.02 ─ 08
Band
Frequency start [GHz]
Frequency stop [GHz]
E
60.0
90.0
W
75.0
110.0
F
90.0
140.0
D
110.0
170.0
G
140.0
220.0
J
220.0
325.0
Y
325.0
500.0
USER
32.18
(default)
68.22
(default)
*)
The band formerly referred to as "A" is now named "KA".
[SENSe:]MIXer<x>:HARMonic:HIGH:STATe
<State>
This command specifies whether a second (high) harmonic is to be used to cover the
band's frequency range.
Suffix:
<x>
.
1..n
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
MIX:HARM:HIGH:STAT ON
Manual operation:
See
[SENSe:]MIXer<x>:HARMonic:HIGH[:VALue]
<HarmOrder>
This command specifies the harmonic order to be used for the high (second) range.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<HarmOrder
numeric value
Range:
3 to 128 (USER band); for other bands: see band
definition
Example:
MIX:HARM:HIGH:STAT ON
MIX:HARM:HIGH 3
Manual operation:
See
Setting basic measurement parameters