
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
938
User Manual 1178.8520.02 ─ 08
Unlike manual operation, you can define a relative limit regardless of the limit check
mode.
Suffix:
<sb>
.
1 to 8
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Level>
Relative limit at the start frequency of a SEM range.
Range:
-400 to 400
*RST:
-50
Default unit: dBc
Example:
SENS:ESP:RANG:LIM:REL:STAR -10
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt:ABS
<Level>
This command defines an absolute limit for the MAX function of the relative limit for a
SEM range.
For more information see
"Relative limit line functions"
Suffix:
<sb>
.
1 to 8
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Level>
Absolute limit at the start frequency of a SEM range to be used
in addition to the relative limit if the MAX function is enabled (see
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:
RELative:STARt:FUNCtion
Range:
-400 to 400
*RST:
-13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STARt:
ABSolute -10
Chapter 12.5.6.11, "Example: SEM
Manual operation:
See
Configuring and performing measurements