Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1145
User Manual 1178.8520.02 ─ 08
<t>
Parameters for setting and query:
<State>
OFF
Switch the function off
ON
Switch the function on
*RST:
OFF
Example:
DISP:WIND2:TRAC1:LAB ON
DISP:WIND2:TRAC1:LAB:TEXT 'MaxTrace'
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:LABel:TEXT
<Text>
Defines a descriptive label for the specified trace instead of the default "Trace <x>"
label.
Enable the label using the
DISPlay[:WINDow<n>]:TRACe<t>:LABel[:STATe]
command.
You can only configure labels for active traces and for traces whose "State" is enabled.
Suffix:
<n>
.
<t>
Parameters:
<Text>
String containing the trace label.
Example:
DISP:WIND2:TRAC1:LAB ON
DISP:WIND2:TRAC1:LAB:TEXT 'MaxTrace'
Manual operation:
See
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:MODE
<Mode>
This command selects the trace mode. If necessary, the selected trace is also activa-
ted.
For max hold, min hold or average trace mode, you can set the number of single mea-
surements with
. Note that synchronization to the end of the
measurement is possible only in single sweep mode.
Suffix:
<n>
.
<w>
subwindow
Not supported by all applications
Analyzing measurements (basics)